RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew Inc.

Recalling firmSmith & Nephew (Suzhou) Limited
Registered facilitySMITH & NEPHEW (SUZHOU) LIMITED (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonReports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo
ProductSmith & Nephew RENASYS-F XL with Soft Port REF #: 66800797 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)
Quantity15,754 indivdual units
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singap…
Initiated2014-01-09
Classified2014-03-10
Terminated2014-09-26
Recall numberZ-1159-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew Inc. (Z-1159-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67477/

Not open yet

Alerts for Smith & Nephew (Suzhou) Limited

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.