Class II — Philips Medical Systems, Inc.

Recalling firmPhilips Medical Systems International BV
Registered facilityView this company's 54 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonHeartStart XL+ battery charge time to 100% capacity at 35C (95F) is slightly longer than the specified 3 hour duration as stated in the labeling
ProductPhilips HeartStart XL+ Defibrillator/Monitor Model: 861290 The HeartStart XL+ is intended for use in a hospital setting by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support or defibrillation. When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart XL+ is suitable for use by medical personnel trained in basic life support that includes the use of an AED. When operating in Monitor, Manual Defibrillation or Pacing modes, the HeartStart XL+ is suitable for use by healthcare pro…
Quantity7157 units
DistributionWorldwide Distribution - USA (nationwide) including Puerto Rico and the countries of ALGERIA ANGOLA ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BELGIUM BOLIVIA BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA (local name: Hrvatska) CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REPUBLIC EGYPT EL SALVADOR ESTONIA ETHIOPIA FINLAND FRANCE GABON GERMANY GREECE GUATEMALA HONG KONG HUNGARY ICELAND I…
Initiated2014-01-24
Classified2014-02-26
Terminated2020-09-03
Recall numberZ-1107-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Philips Medical Systems, Inc. (Z-1107-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67328/

Not open yet

Alerts for Philips Medical Systems International BV

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.