Class II — Siemens Healthcare Diagnostics
| Recalling firm | SIEMENS HEALTHCARE DIAGNOSTICS INC. |
|---|---|
| Registered facility | View this company's 12 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays. |
| Product | IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum. |
| Quantity | 18,489 units |
| Distribution | Worldwide distribution. |
| Initiated | 2013-12-20 |
| Classified | 2014-04-25 |
| Terminated | 2018-05-18 |
| Recall number | Z-1518-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Siemens Healthcare Diagnostics (Z-1518-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67210/