Class II — Carestream Health Inc.
| Recalling firm | Carestream Health, Inc. |
|---|---|
| Registered facility | View this company's 6 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor. |
| Product | Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures. |
| Quantity | 4 units |
| Distribution | Nationwide Distribution including TX, IL, and GA. |
| Initiated | 2013-08-20 |
| Classified | 2014-01-29 |
| Terminated | 2016-06-28 |
| Recall number | Z-0774-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Carestream Health Inc. (Z-0774-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/67060/
Alerts for Carestream Health, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.