Class II — Draeger Medical Systems, Inc.

Recalling firmDraeger Medical Systems, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonAfter a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.
ProductDraeger Infinity Acute Care System Monitoring Solution
Quantity439
DistributionUS Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
Initiated2013-12-10
Classified2014-01-10
Terminated2015-05-20
Recall numberZ-0714-2014
InitiationVoluntary: Firm initiated