Class III — Bio-Rad Laboratories, Inc.

Recalling firmBio-Rad Laboratories, Inc.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass III
StatusTerminated
ReasonD-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.
ProductD-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.
Quantity2906 kits (318 Kits in US; 2588 kits internationally)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.
Initiated2013-11-22
Classified2014-02-10
Terminated2014-03-20
Recall numberZ-0965-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class III recall — Bio-Rad Laboratories, Inc. (Z-0965-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66947/

Not open yet

Alerts for Bio-Rad Laboratories, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.