Class II — CareFusion 303, Inc.
| Recalling firm | CareFusion 303, INC. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | CareFusion is recalling the VersaSafe Extension Intravascular Administration Set because it may have excessive adhesive which will prevent the male spin collar to operate as intended. |
| Product | VersaSafe Extension Set, Model No. 21000M-007 |
| Quantity | 3500 units |
| Distribution | Nationwide Distribution. |
| Initiated | 2013-11-08 |
| Classified | 2013-12-03 |
| Terminated | 2014-04-17 |
| Recall number | Z-0435-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — CareFusion 303, Inc. (Z-0435-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66758/
Alerts for CareFusion 303, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.