Class II — Mindray DS USA, Inc. dba Mindray North America
| Recalling firm | Mindray DS USA, Inc. |
|---|---|
| Registered facility | MINDRAY DS USA, Inc. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue. |
| Product | Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Product Usage: The V Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use for the V Series include the monitoring of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements 2) Heart Rate 3) Pulse Oximtery 4) ST Segment 5) Arrhythmia Detection 6) Non Invasive Blood Pressure (IBP) 7) Cardiac Output (CO) 8) Respiratory Gases 9) Respiration Rate 10) Temperature The V series monitor h… |
| Quantity | 1839 units |
| Distribution | Worldwide Distribution - USA Nationwide |
| Initiated | 2013-10-07 |
| Classified | 2013-11-25 |
| Terminated | 2015-04-27 |
| Recall number | Z-0389-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Mindray DS USA, Inc. dba Mindray North America (Z-0389-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66755/
Alerts for Mindray DS USA, Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.