Class II — GE Healthcare, LLC
| Recalling firm | GE HealthCare Technologies Inc. |
|---|---|
| Registered facility | View this company's 50 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | GE Healthcare has recently become aware of a potential safety issue due to the image orientation within GSI viewer on AW products. The GSI Viewer application running on AW/Volume Share or on the AW Server when making a batch and rotating the batch group clockwise 180 degrees (Inferior to Superior) will display images in the Superior orientation. The orientation annotation on the image is correc |
| Product | AW VolumeShare 4 (ADVANTAGE WORKSTATION 4.3.), AW VolumeShare 5 (AW VOLUMESHARE 5 WITH ANGIOVIZ OPTION), AW Server. K052995: Advantage Workstation 4.3 is a review station, which allows easy selection, review, processing and filming of multi-modality DICOM images from a variety of diagnostic imaging systems. When interpreted by a trained physician, filmed or displayed images on the AW monitor may be used as a basis for diagnosis, except in the case of mammography images. K110834: AW VolumeShare 5 is a review workstation, which allows easy selection, review, processing and filming of multi-… |
| Quantity | 1458 (336 US, 1122 OUS) |
| Distribution | Worldwide distribution: US (nationwide) including states of: AL, AZ, CA, CT,DE, FL,GA, HI, IL, IN, IA, KS, MD,MA, MI,MN, MS, O. NV, NH, NJ, NY, NC, OH, OK, OR, PA, SD, TN, TX , UT, VA, WA, WI. OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, BYELO RUSSIAN, CANADA, CHINA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONDURAS, HON KONG, HUNGARY, INDIA, INDONESIA, IRAQ, IRELAND, I… |
| Initiated | 2013-09-19 |
| Classified | 2013-11-13 |
| Terminated | 2014-07-31 |
| Recall number | Z-0270-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — GE Healthcare, LLC (Z-0270-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66550/
Alerts for GE HealthCare Technologies Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.