Class II — Stryker Endoscopy
| Recalling firm | STRYKER CORP. |
|---|---|
| Registered facility | View this company's 61 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | When positioning the display located on top of the video cart, it is possible for the user to apply excessive force and push the display up and out of the mounting arm. This can result in the display falling upon the user/patient. |
| Product | Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients. Model number 0240-095-200, manufactured by Stryker Endoscopy San Jose The intended use of the Stryker Vision Mount is to hold and position an LCD monitor. The Stryker Vision Mounting Arm can be used with any Stryker flat panel display and is It is compatible with Stryker Carts (0240099001, 0240099011 or 0240099020). It is indicated for use in a surgical operating room environment and is exposed to environmental conditions associated with the storage and shipping of the product. It … |
| Quantity | 14271 units |
| Distribution | Worldwide Distribution - USA including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and internationally to Australia, Canada, Chile, China, Argentina, Guam, Taiwan, India, Japan, Korea, Latin America, Philippines, Mexico, New Zealand, Hong Kong… |
| Initiated | 2013-08-30 |
| Classified | 2013-09-24 |
| Terminated | 2014-04-30 |
| Recall number | Z-2271-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Stryker Endoscopy (Z-2271-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66211/
Alerts for STRYKER CORP.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.