Class II — DePuy Mitek, Inc., a Johnson & Johnson Co.

Recalling firmDEPUY MITEK, A JOHNSON & JOHNSON COMPANY
Registered facilityDePuy Mitek, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonPillow valve included in the FMS Duo outflow tubing set and the FMS Solo intermediary tubing set may not perform as intended. may result in backflow of irrigation fluid into the One Day Set which may potentially lead to patient cross-contamination
ProductDePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS® Fluid Management Systems connects the Inflow Tubing to the arthroscopic sheath during arthroscopic procedures. The Polyvinyl Chloride (PVC) tubing integrates several functional components including: luer locks for attaching the tube to the Inflow tubing and sheath, clamps, and a one-way valve. The FMS Intermediary Tubing is sterile and intended for single use with the FMS Fluid Management Systems in a surgical setting by personnel trained in arthroscopy .
Quantity368 pk
DistributionWorldwide Distribution - USA (Nationwide) and in the countries of: Australia , Austria , Belgium , Canada , Czech Republic , Denmark , Finland , France, Germany , India, Ireland , Israel, Italy , Latvia , Netherlands , Norway , Nepal, Poland , Portugal , Spain , Sweden , Switzerland , United Kingdom.
Initiated2013-08-29
Classified2013-11-08
Terminated2016-05-12
Recall numberZ-0191-2014
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — DePuy Mitek, Inc., a Johnson & Johnson Co. (Z-0191-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66145/

Not open yet

Alerts for DEPUY MITEK, A JOHNSON & JOHNSON COMPANY

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.