Class II — GE Healthcare, LLC
| Recalling firm | GE HealthCare Technologies Inc. |
|---|---|
| Registered facility | View this company's 50 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | When performing a secondary capture generated from GSI Viewer, the header information is correct on the images. However, when changing the ROI (region of interest) and zooming in, the DICOM header does not change to reflect the new image, thus the line measurements may be inaccurate. |
| Product | GE Healthcare, Model Number/ Description: 5342114 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-2 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-3 GSI Viewer 1.10 on AW VolumeShare 4 (AW 4.5); 5342114-4 GSI Viewer 1.11 on AW VolumeShare 4 (AW 4.5); 5401757 GSI Viewer 1.20 on AW VolumeShare 3 (AW4.4). Computed Tomography X-Ray system |
| Quantity | 283 (59 US, 224 OUS) |
| Distribution | Worldwide Distribution: US (nationwide): AL, AZ, CA, CT, FL, IL, IN, IA, KY, MD, MA, MI, MS, NV, NH, NJ, NY, PA, WA, and WI; and countries of: AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, KUWAIT, LATVIA, NEW ZEALAND, POLAND, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SOUTH KOREA, SWEDEN,, SWITZERLAND, TAIWAN, a… |
| Initiated | 2011-03-25 |
| Classified | 2014-03-06 |
| Terminated | 2014-03-13 |
| Recall number | Z-1148-2014 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — GE Healthcare, LLC (Z-1148-2014).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/66119/
Alerts for GE HealthCare Technologies Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.