RegWardenFDA facility compliance records
Recall

Class II — DeRoyal Industries Inc

Recalling firmDEROYAL INDUSTRIES, INC.
Registered facilityView this company's 13 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonDeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.
ProductDeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.
Quantity506,632 sets in US; 17,854 sets International
DistributionWorldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Dubai, Arab Emirates, Quebec, Canada, Japan, and Ireland.
Initiated2013-07-15
Classified2013-08-28
Terminated2015-05-20
Recall numberZ-2113-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — DeRoyal Industries Inc (Z-2113-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65764/