Class II — Carestream Health, Inc.
| Recalling firm | Carestream Health, Inc. |
|---|---|
| Registered facility | View this company's 6 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch. |
| Product | Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images. |
| Quantity | 180 units |
| Distribution | Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN. |
| Initiated | 2013-06-06 |
| Classified | 2013-10-31 |
| Terminated | 2016-05-20 |
| Recall number | Z-0126-2014 |
| Initiation | Voluntary: Firm initiated |