Class II — Applied Medical Resources Corp

Recalling firmAPPLIED MEDICAL RESOURCES
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonApplied Medical is conducting a voluntary recall on specific lot numbers of its Epix® and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
ProductApplied Medicals Epix® and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
Quantity127,305 units
DistributionWorldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Ta…
Initiated2013-05-01
Classified2013-06-07
Terminated2014-02-06
Recall numberZ-1521-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Applied Medical Resources Corp (Z-1521-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65217/

Not open yet

Alerts for APPLIED MEDICAL RESOURCES

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.