RegWardenFDA facility compliance records
Recall

Class II — GE Healthcare, LLC

Recalling firmGE HealthCare Technologies Inc.
Registered facilityView this company's 50 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonScreen save images are included in wrong exam due to duplicate UID creation. This may cause a safety issue due to a potential patient data mismatch.While the issue has only been observed at one customer site on the Signa HDxT 1.5T system, internal investigation revealed that the issue also impacts multiple GE healthcare MR Systems.
ProductSIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM GE SIGNA 3.0T WITH EXCITE MR SYSTEM GE SIGNA EXCITE 1.5T MR SYSTEM GE SIGNA EXCITE 3.0T MR SYSTEM Signa HDx Signa HDxt Signa HDi Signa Vibrant GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (i.e. Signa Ovation 5 and Ovation HD) GE 1.5T SIGNA HDe MR SYSTEM GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM GE 0.7T SIGNA OPENSPEED (4) WITH EXCITE MR SYSTEM GE 0.7T SIGNA OPENSPEED (5) WITH EXCITE MR SYSTEM The GE Signa Infinity line of products are whole body magnetic resonance…
Quantity7016 (2919 US; 4097 OUS)
DistributionWorldwide distribution: USA (nationwide) including DC and Puerto Rico and countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIDJAN, BELARUS, BELGIUM, BOLIVIA, BOSNIA-HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, FINLAND, FRANCE, GEORGIA, GERMANY, GREAT BRITAIN, GREECE, GUADELOUPE, GUATEM…
Initiated2011-11-21
Classified2013-08-09
Terminated2014-10-22
Recall numberZ-1943-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GE Healthcare, LLC (Z-1943-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65133/

Not open yet

Alerts for GE HealthCare Technologies Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.