RegWardenFDA facility compliance records
Recall

Class II — Fresenius Medical Care Holdings, Inc. dba Renal Solutions

Recalling firmFresenius Medical Care Holdings, Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonProduct is being recalled because the combination of the current 50 mg vial, the 2008 MeDS pump, and the 2008 MeDS pump tubing Set for CombiSet Bloodlines can lead to a situation where Venofer(R) delivery does not meet dosage accuracy requirements. This may lead to a slight over or under delivery of Venofer(R) that does not pose an acceptable risk to the patient. Please note that this recall
Product2008 MeDS Pump, Venofer(R) Pump Product Usage: The 2008 MeDs pump is an optional accessory for use with the Fresenius 2008 Series Hemodialysis Machines and is intended to administer Venofer to treat iron deficiency anemia in patients with chronic kidney disease undergoing hemodialysis, where Venofer, in conjunction with hemodialysis, is prescribed by a physician.
Quantity132
DistributionUSA Nationwide Distribution in the states of CA, MA, and NY.
Initiated2013-04-01
Classified2013-08-26
Terminated2015-03-02
Recall numberZ-2082-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Fresenius Medical Care Holdings, Inc. dba Renal Solutions (Z-2082-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/65080/

Not open yet

Alerts for Fresenius Medical Care Holdings, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.