Class II — Elekta, Inc.
| Recalling firm | ELEKTA AB |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The Column rotation can accidentally be moved to a different position which in the locked position. |
| Product | Precise Treatment Table. To be used as part of radiation therapy treatment process. |
| Quantity | 443 |
| Distribution | Nationwide Distribution and Puerto Rico. |
| Initiated | 2013-03-14 |
| Classified | 2013-04-09 |
| Terminated | 2015-03-17 |
| Recall number | Z-1072-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Elekta, Inc. (Z-1072-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64664/
Alerts for ELEKTA AB
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.