Class II — Roche Diagnostics Operations, Inc.

Recalling firmRoche Diagnostics Operations Inc.
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonRoche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
ProductCoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Quantity4467 both devices
DistributionNationwide Distribution.
Initiated2013-03-06
Classified2013-04-15
Terminated2018-01-10
Recall numberZ-1114-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Roche Diagnostics Operations, Inc. (Z-1114-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64460/

Not open yet

Alerts for Roche Diagnostics Operations Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.