RegWardenFDA facility compliance records
Recall

Class II — Varian Medical Systems, Inc. Oncology Systems

Recalling firmVarian Medical Systems, Inc
Registered facilityView this company's 14 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThe calculation of dose dynamic plans containing different energy fields is incorrect when the Pencil Beam Convolution 11.0 algorithm (PBC 11.0) is used.
ProductVarian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Quantity1168 units
DistributionWorldwide Distribution - USA Nationwide
Initiated2013-01-28
Classified2013-02-20
Terminated2014-04-04
Recall numberZ-0867-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Varian Medical Systems, Inc. Oncology Systems (Z-0867-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64319/