Class II — Mindray DS USA, Inc. d.b.a. Mindray North America

Recalling firmMindray DS USA, Inc.
Registered facilityMINDRAY DS USA, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonMindray has identified a software anomaly that may where, in the presence of an asystole, the Telepack ECG waveform may produce a unique step pattern (instead of a flat line) if Lead 1 or V is being utilized for detecting cardiac arrhythmia.
ProductMindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, and 300BF-PA00003 (DPM Central Monitoring System) 115-006908-01, -03, -04 (DPM software upgrade kit). The DPM Central Monitoring System transfers information between DPM Central station and other networked devices.
Quantity51 units
DistributionNationwide Distribution.
Initiated2012-11-28
Classified2013-02-11
Terminated2013-08-15
Recall numberZ-0798-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Mindray DS USA, Inc. d.b.a. Mindray North America (Z-0798-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/64114/

Not open yet

Alerts for Mindray DS USA, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.