Class II — Integra LifeSciences Corporation
| Recalling firm | Integra LifeSciences Corporation |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | An investigation of an adverse trend of complaints conducted by Integra LifeSciences identified various product performance issues which could impact the functionality of the firm's Mayfield Composite Series Base Units and Mayfield Skull Clamps. In order to prevent additional complaint reports, the firm decided to recall and replace certain lot codes of the Base Units and Skull Clamps that were d |
| Product | MAYFIELD Composite Series Skull Clamps, Model A3059, 510(k) #K120633. The MAYFIELD Skull Clamps (A3059) are indicated for use in open and percutaneous craniotomies, as well as spinal surgery when rigid skeletal fixation is necessary. |
| Quantity | eight (8) skull clamps were distributed |
| Distribution | Nationwide Distribution including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NY, OH, OK, OR, PA TN, TX, UT, VA, WA and WI. |
| Initiated | 2012-11-05 |
| Classified | 2012-12-19 |
| Terminated | 2014-02-21 |
| Recall number | Z-0559-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Integra LifeSciences Corporation (Z-0559-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63776/
Alerts for Integra LifeSciences Corporation
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.