Class II — GE Healthcare, LLC
| Recalling firm | GE HealthCare Technologies Inc. |
|---|---|
| Registered facility | View this company's 50 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | GE has become aware of multiple issues affecting the Dash 3000/4000/5000 Patient Monitor. These issues may occur when the Ethernet controller in the XPC823 CPU on the Dash and DashPort 2 motherboard functions intermittently. GE Healthcare has identified that these issues occur most often at sites with large Unity Network installations. It should be noted that the Dash 3000/4000/5000 patient m |
| Product | GE Healthcare, Dash 3000, Dash 4000 and Dash 5000 Physiological Patient Monitor and accessories. |
| Quantity | 119,262 |
| Distribution | Nationwide Distribution - ALL STATES IN CONTINENTAL USA INCLUDING PR and DC excluding AK, DE, ND, and VT. ALGERIA, AUSTRALIA, ANGOLA, ARGENTINA, AUSTRIA, BAHAMAS, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CUBA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRAN… |
| Initiated | 2012-06-13 |
| Classified | 2013-02-13 |
| Terminated | 2016-08-25 |
| Recall number | Z-0813-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — GE Healthcare, LLC (Z-0813-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63654/
Alerts for GE HealthCare Technologies Inc.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.