Recall

Class II — King Systems Corp.

Recalling firmKING SYSTEMS
Registered facilityKING SYSTEMS CORPORATION. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
ProductKING LAD Standard ClearSeal SINGLE USE, King Systems ETO Sterilie, 10 devices/case Size 5 / Adult ClearSeal King LAD® Patient Size: >70kg Product Usage: Laryngeal airway
Quantity263 cases of product. (2,630 devices)
DistributionUS Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
Initiated2012-11-02
Classified2012-11-21
Terminated2013-05-28
Recall numberZ-0424-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — King Systems Corp. (Z-0424-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63580/

Not open yet

Alerts for KING SYSTEMS

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.