Class II — Zimmer, Inc.
| Recalling firm | ZIMMER, INC. |
|---|---|
| Registered facility | View this company's 10 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Zimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the inserter instrument was inadvertently removed from patient with the implant still engaged. The correction includes revised directions explaining the technique to remove the implant |
| Product | The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis Implant into the cleared disc space. The implant is secured to the inserter via finger-tightening the knob of the inserter shaft allowing the lateral grasping arms to secure the implant. Once the implant is inserted and its position confirmed radiographically, the inserter is removed from the implant by turning the shaft knob counterclockwise to loosen the lateral grasping arms from the implant. |
| Quantity | 170 |
| Distribution | USA nationwide; Europe, Sweden |
| Initiated | 2012-09-20 |
| Classified | 2013-01-09 |
| Terminated | 2014-04-23 |
| Recall number | Z-0659-2013 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Zimmer, Inc. (Z-0659-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63559/
Alerts for ZIMMER, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.