RegWardenFDA facility compliance records
Recall

Class II — Encore Medical, Lp

Recalling firmENCORE MEDICAL, L.P.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on
Product6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.
Quantity1 unit
DistributionNationwide Distribution only to CA.
Initiated2012-09-28
Classified2012-10-25
Terminated2013-01-25
Recall numberZ-0139-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Encore Medical, Lp (Z-0139-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63321/