Class II — Beckman Coulter Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBeckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar
ProductUniCel® DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel® DxC Synchron® Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Quantity7 units total (both US and International)
DistributionWorldwide Distribution - USA including TX and OH. Interntionally to China and Turkey
Initiated2012-09-03
Classified2012-10-24
Terminated2014-02-14
Recall numberZ-0123-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter Inc. (Z-0123-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/63143/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.