RegWardenFDA facility compliance records
Recall

Class II — Orthofix, Inc

Recalling firmORTHOFIX SRL
Registered facilityORTHOFIX SRL (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is a possibility that the ISKD limb lengthener may stop distracting post-operatively during treatment, which may result in premature bone consolidation (limb not achieving the desired length) leading to revision surgery to remove and/or replace the device.
ProductOrthofix ISKD Limb Lengthener. ISKD - Intramedullary Skeletal Kinetic Distractor. The ISKD is intended for limb lengthening of the tibia and femur.
Quantity1224
DistributionWorldwide Distribution - USA (nationwide) Internationally to Australia, Hong Kong, Saudi Arabia, Korea, Turkey, Greece, Finland, South Africa, Spain, Singapore, Switzerland, Brazil, Germany, France, Japan, Netherlands, New Zealand, Israel, Canada and Mexico
Initiated2012-08-06
Classified2013-02-28
Terminated2013-05-17
Recall numberZ-0895-2013
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Orthofix, Inc (Z-0895-2013).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62804/