Class II — Siemens Healthcare Diagnostics, Inc.

Recalling firmSIEMENS HEALTHCARE DIAGNOSTICS INC.
Registered facilityView this company's 12 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhen using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall notification. QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient results will not be held as expecte
Productsyngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments.
Quantity44
DistributionUSA ( nationwide ) including the states of AZ, CA, FL, IL, KY, MD, MA, MN, MS, MO, NE, NH, NJ, NY, NC, OH, PA, TX, VA and WV.
Initiated2012-05-29
Classified2012-08-31
Terminated2014-04-10
Recall numberZ-2305-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Healthcare Diagnostics, Inc. (Z-2305-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62709/

Not open yet

Alerts for SIEMENS HEALTHCARE DIAGNOSTICS INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.