RegWardenFDA facility compliance records
Recall

Class II — Terumo Cardiovascular Systems Corporation

Recalling firmTerumo Cardiovascular Systems Corp.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonTerumo Cardiovascular Systems has received reports of malfunctions involving the loss of data transfer capabilities using the data transfer capabilities using modules on the Terumo Advanced Perfusion System 1. Some reports stated that the odor of smoke was as associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the modu
ProductSerial interface module RS-485 Product Usage: The System 1 provides perfusion case data transfer through the use of the data transfer module. This module is a general purpose serial interface and is the interface between the TLink Data Management ystem or Sarns CCPro Data Perfusion oftware and the system. The data collected and exported by the data transfer module includes physiological data and device events to record the progress and events during a procedure.
Quantity34 units
DistributionWorldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, and WV and the countries of: AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Germany, Hong Kong, Indonesia, Japan, Jordan, Malaysia, Singapore, South Af…
Initiated2012-07-26
Classified2012-08-20
Terminated2014-06-27
Recall numberZ-2231-2012
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Terumo Cardiovascular Systems Corporation (Z-2231-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/62684/

Not open yet

Alerts for Terumo Cardiovascular Systems Corp.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.