Class II — Polymer Technology Systems, Inc.
| Recalling firm | POLYMER TECHNOLOGY SYSTEMS, INC. |
|---|---|
| Registered facility | Polymer Technology Systems (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not the reportable measuring range for the CardioChek PA meter for the creatinine assay. |
| Product | PTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips are intended to measure creatinine in whole blood. Creatinine measurements are used in the diagnosis and treatment of renal (kidney) diseases and in the monitoring of renal dialysis. This system is intended for the professional use for accurately and reproducibly measuring creatinine in human whole blood obtained from a finger stick sample |
| Quantity | 1650 vials |
| Distribution | Worldwide Distribution -- Nationwide Distribution including the states of CA and VA., and the countries of China, Poland, Ireland and France. |
| Initiated | 2012-05-15 |
| Classified | 2012-06-28 |
| Terminated | 2012-07-20 |
| Recall number | Z-1903-2012 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Polymer Technology Systems, Inc. (Z-1903-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/61966/
Alerts for POLYMER TECHNOLOGY SYSTEMS, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.