Class II — Etac Supply Center Ab
| Recalling firm | Etac Supply Center AB |
|---|---|
| Registered facility | ETAC SUPPLY CENTER AB (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The recall has been initiated because there is the potential that users may fall and become injured, with the potential for serious injury, including bone fracture. One death occurred in a user fall incident, but it is unclear whether the fall was caused by the loss of a wheel. Among the walkers distributed worldwide, Etac® has received a total of 49 reports from Europe and Japan of wheels falling |
| Product | Etac® Ono Walker, Etac® Avant Walker, and Etac® Salsa Walker. These products are four-wheeled walkers intended as a walking aid for indoor and outdoor use for persons weighing up to 125 KG. |
| Quantity | 204,323 units of the Etac Ono Walker, 15,601 units of the Etac Avant Walker, and 6,771 units of the ETAC Salsa Walker |
| Distribution | Worldwide Distribution-USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, China, Colombia, Germany, Denmark, Estonia, Spain, Finland, France, Great Brittain, Greenland, Greece, Hungaria, Ireland, Israel, Iceland, Italy, Japan, South Korea, Kuwait, Lithuania, Luxembourg, Latvia, Malaysia, Netherlands, Norway, Poland, Portugal, Rumania, Rus… |
| Initiated | 2012-04-06 |
| Classified | 2012-07-14 |
| Terminated | 2013-06-18 |
| Recall number | Z-2014-2012 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Etac Supply Center Ab (Z-2014-2012).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/61463/
Alerts for Etac Supply Center AB
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.