Class II — Terumo Cardiovascular Systems Corp
| Recalling firm | Terumo Cardiovascular Systems Corp. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Defect in the luer connector can compromise the seal on the female side of the connector. and the compromised seal can introduce air in the table line |
| Product | Terumo Custom Cardioplegia Delivery Set- Catalog Number:165720 and Catalog Number: 6375 Product Usage: Cardioplegia Delivery System that is designed to delivery blood/solution to the patient undergoing cardiopulmonary bypass surgery. The set is designed to deliver blood/solutions at a 4: I ratio. Includes a Conducer Heat Exchanger and a MP4 Temperature & Pressure Module. Additionally, the set is constructed with a female luer connector that is the subject of the correction/removal activity. |
| Quantity | Catalog: 16579 ; Catalog 6375: 3256 units |
| Distribution | Worldwide Distribution: US (nationwide) and the countries of Canada, Brazil, Hong Kong, Korea, and Thailand. |
| Initiated | 2011-03-01 |
| Classified | 2014-10-31 |
| Terminated | 2014-11-06 |
| Recall number | Z-0160-2015 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Terumo Cardiovascular Systems Corp (Z-0160-2015).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/58185/