RegWardenFDA facility compliance records
Facility

OriGen Biomedical, Inc.

9 recall(s) on record — most severe on record: Class I.

Activity by year

FDA actions on record by year2015: 2 recalls; 2017: 2 recalls; 2018: 2 recalls; 2019: 1 recalls; 2021: 1 recalls; 2023: 1 recalls220152016220172201812019202012021202212023202420252026
  • Recalls (9)

Devices registered here (1)

Recalls & enforcement

  • Class IIOngoing2023-12-19
    Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resul…
  • Class IIITerminated2021-01-28
    It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
  • Class IITerminated2019-06-24
    The firm has become aware that they have distributed product that was expired.
  • Class IITerminated2018-08-21
    These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
  • Class IITerminated2018-08-21
    The catheter failed the endotoxin testing.
  • Class ITerminated2017-08-02
    Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO Catheter has had a separation of the clear extension tube from the hub connection. These adverse events resulted in patient injury (blood loss).
  • Class ITerminated2017-01-17
    Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later f…
  • Class ITerminated2015-03-30
    Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage.
  • Class ITerminated2015-03-30
    Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “OriGen Biomedical, Inc. (FEI 3042484745) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3042484745-origen-biomedical-inc/

Not open yet

Alerts for OriGen Biomedical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.