SUNMED GROUP HOLDINGS, LLC dba AirLife
6 recall(s) on record — most severe on record: Class I.
Activity by year
- Recalls (6)
Devices registered here (1)
- Tube, Aspirating, Flexible, ConnectingProduct code BYY (opens in a new tab) · Class II · General Hospital · 21 CFR 880.6740
Recalls & enforcement
- Class IOngoing2025-12-15Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
- Class IOngoing2025-05-15The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant h…
- Class IOngoing2025-05-01Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequence…
- Class IIOngoing2023-10-11A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
- Class IIOngoing2022-12-22There is an orogastric (OG) tube size discrepancy between the labeling in the IFU and the printed text on the device. If the clinician is unable to pass a gastric catheter through one of the channels, the clinician may use a smaller-diameter gastric catheter.
- Class IITerminated2022-09-08The kits are being recalled due to a lack of sterility of a spare needle contained within the kits. There is a potential for skin infection or sepsis if the patient is exposed to a non-sterilized, unused needle.
Related
- All facilities operated by SunMed Group Holdings, LLC dba AirLife
- Westmed de Mexico, S.A de C.V (MX)
- PRODUCTOS UROLOGOS DE MEXICO S.A. DE C.V. (MX)
- SunMed Group Holdings, LLC (MI)
- AirLife Finland Oy (FI)
- Westmed de Mexico, S.A de C.V (MX)
- Shenzhen Vital Signs-KTL Medical Instrument Co Ltd (CN)
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “SUNMED GROUP HOLDINGS, LLC dba AirLife (FEI 3030308442) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3030308442-sunmed-group-holdings-llc-dba-airlife/
Alerts for SUNMED GROUP HOLDINGS, LLC dba AirLife
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.