Baja Fur S.A. de C.V.

5 recall(s) on record — most severe on record: Class I.

Activity by year

FDA actions on record by year2022: 1 recalls; 2023: 1 recalls; 2024: 3 recalls2017201820192020202112022120233202420252026
  • Recalls (5)

Devices registered here (1)

  • Set, Tubing, Blood, With And Without Anti-Regurgitation Valve
    Product code FJK (opens in a new tab) · Class II · Gastroenterology, Urology · 21 CFR 876.5820

Recalls & enforcement

  • Class IIOngoing2024-08-01
    Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experien…
  • Class IIOngoing2024-04-30
    A cart with prefiltration drawer, an optional hemodialysis system accessory, has a cord with a female connector that has internal contact pins that may become damaged that could create localized heat, sparks, smoke, melting or charring (thermal event) of the female connector or t…
  • Class IOngoing2024-03-06
    Hemodialysis System, pre and post dialyzer peroxide-cured silicone tubing, with less than 336 hours usage, will be replaced because tubing may contain non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs), that have not been flushed out over time with usage. Exposure to PC…
  • Class IIOngoing2023-01-24
    An observed trend of high conductivity dialysate alarms in a hemodialysis system as a result of introducing a software tool into the manufacturing process.
  • Class IIOngoing2022-01-20
    Due to a component in the hemodialysis console there is the possibility of heat-related damage that may occur with the device.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Baja Fur S.A. de C.V. (FEI 3017552925) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3017552925-baja-fur-s-a-de-c-v/

Not open yet

Alerts for Baja Fur S.A. de C.V.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.