ALPHATEC SPINE, INC.
17 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (17)
Devices registered here (1)
- Orthopedic Stereotaxic InstrumentProduct code OLO (opens in a new tab) · Class II · Neurology · 21 CFR 882.4560
Recalls & enforcement
20262 records
- Class IIOngoing2026-05-13Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication between the Patient Interface and Tablet, and prompts requiring sy…
- Class IIOngoing2026-01-02Due a design issue where the navigated array connection geometry is incorrect.
20242 records
- Class IIOngoing2024-12-15Complaints have been received regarding post-operative implant collapse.
- Class IIOngoing2024-01-26Due to reports of intraoperative graft bolt implantation breakages.
20231 record
- Class IIOngoing2023-07-07Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to ove…
20221 record
- Class IIOngoing2022-09-14Due to manufacturing instruments which included additional features and adequate consideration was not given to the need for a new Regulatory filing.
20211 record
- Class IITerminated2021-04-26There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
20171 record
- Class IITerminated2017-05-05Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.
20161 record
- Class IITerminated2016-10-05Alphatec Spine is recalling the Arsenal Spinal Fixation System Set Screw due to a trend in set screw postoperative disengagement from the screw body.
20151 record
- Class IITerminated2015-04-10During a field audit it was discovered by an employee that a 16mm Trestle Luxe Screw has been both laser marked and labeled as a 14mm Trestle Luxe Screw (4.0mm Variable Angle, Self-Drilling Hexalobe Screw, 14mm). As a result the firm recalling the affected lots codes for this mi…
20141 record
- Class IITerminated2014-06-26Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.
20132 records
- Class IITerminated2013-06-28The firm initiated this recall of certain lots of Illico Ti Cannulated Polyaxial Screws 7.5mm x 40mm implants, due to an off-center cannulation hole at the proximal end of the bone screw. There is a potential risk that when the screw is mated to a polyaxial screwdriver, the guide…
- Class IITerminated2013-05-06Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head …
20122 records
- Class IITerminated2012-09-25The firm initiated the recall because the stainless steel guide-wire of the Illico MIS Fixation System did not completely fit/pass through the tip of certain Illico Ti Cannulated Polyaxial Screws, therefore rendering such screws nonfunctional.
- Class IITerminated2012-05-29On May 9th 2012 Alphatec Spine received a complaint which reported that the distal tip of the Trestle Luxe Slide Alignment Tool (Part Number 71721), does not properly fit into the hexalobe screw on the Trestle Luxe Plate.
20111 record
- Class IITerminated2011-07-06The rods were incorrectly laser etched as ILLICO® Straight CP Tl Rods, 110mm, when they were in fact manufactured to 100mm in length.
20101 record
- Class IITerminated2010-12-09The recall was initiated because Alphatec Spine confirmed that the TI Standard Polyaxial Screw Assembly was laser marked with the incorrect length of 45 MM.
Related
- All facilities operated by ALPHATEC HOLDINGS, INC.
- Alphatec Spine, Inc. (TN)
- ALPHATEC SPINE, INC. (CA)
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “ALPHATEC SPINE, INC. (FEI 3015537048) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3015537048-alphatec-spine-inc/
Alerts for ALPHATEC SPINE, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.