PATH
2 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (2)
- Inspections (1)
Devices registered here (3)
- AudiometerProduct code EWO (opens in a new tab) · Class II · Ear, Nose, Throat · 21 CFR 874.1050
- Tester, Auditory ImpedanceProduct code ETY (opens in a new tab) · Class II · Ear, Nose, Throat · 21 CFR 874.1090
- Stimulator, Auditory, Evoked ResponseProduct code GWJ (opens in a new tab) · Class II · Neurology · 21 CFR 882.1900
Recalls & enforcement
- Class IIOngoing2025-11-06Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.
- Class IIOngoing2023-09-26Affected lots of product may experience technical distortions to a greater extent than expected.
FDA inspections (1)
- NAI2017-05-17No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “PATH (FEI 3008111228) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3008111228-path/
Alerts for PATH
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.