Extremity Medical, LLC
4 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (4)
- Inspections (2)
Devices registered here (1)
- Prosthesis, Wrist, Semi-ConstrainedProduct code KWM (opens in a new tab) · Class II · Orthopedic · 21 CFR 888.3800
Recalls & enforcement
- Class IITerminated2013-01-10The 1.6 mm Guidewire is undersized up to 2 mm in length. When used with the depth gauge, the measurement reading will not correspond with the desired screw length. The selected screw size based on the depth gauge reading will be about 2 mm longer than intended.
- Class IITerminated2012-03-22Lag screw measurement is short (approximately 5 mm) when using current screw measuring device (p/n 101-00010) per the IO FIX Surgical Technique Rev C. which can lead to possible inappropriate implant size selection (e.g. shorter screw size selected).
- Class IITerminated2011-04-25Product Event Reports revealed a broken reamer experienced intra-operatively. Pre-drilling was not performed prior to reaming.
- Class IITerminated2011-04-13Product Event Reports revealed that lag screw passes through washer (e.g. X-Post) intra-operatively.
FDA inspections (2)
- NAI2026-06-12No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices
- NAI2016-05-24No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Extremity Medical, LLC (FEI 3007289093) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3007289093-extremity-medical-llc/
Alerts for Extremity Medical, LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.