Zyno Medical LLC.
9 recall(s) on record — most severe on record: Class I.
Activity by year
- Recalls (9)
- Inspections (5)
- Warning letters (2)
Devices registered here (1)
- Set, Administration, IntravascularProduct code FPA (opens in a new tab) · Class II · General Hospital · 21 CFR 880.5440
Recalls & enforcement
- Class IOngoing2025-05-07Unreleased software versions were installed on distributed devices without verification or validation.
- Class IIOngoing2024-10-18Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- Class IIIOngoing2024-09-30For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- Class IOngoing2024-09-13There is a defect in the air-in-line software algorithm.
- Class IIOngoing2024-06-04Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unp…
- Class IIOngoing2023-12-20When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure …
- Class IITerminated2019-05-29Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be dr…
- Class IITerminated2016-07-29Administration Set potential filter leaking
- Class IITerminated2011-06-13Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump
FDA inspections (5)
- VAI2025-09-18Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
- VAI2023-08-16Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
- OAI2022-02-15Official Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
- VAI2017-12-15Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
- OAI2016-08-11Official Action Indicated · Compliance: Devices · citations posted
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
FDA warning letters (2)
- Warning Letter2022-07-12Warning Letter (opens in a new tab) · Natick Massachusetts
- Warning Letter2016-12-05Warning Letter (opens in a new tab) · Natick Massachusetts
Source: U.S. FDA Compliance Actions Dashboard (opens in a new tab) (device warning letters, 2016–present). Data as of 2026-10-06. Each link opens an FDA search for the letter; FDA can take several weeks to post a letter, so a very recent one may not appear yet.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Zyno Medical LLC. (FEI 3006575795) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3006575795-zyno-medical-llc/
Alerts for Zyno Medical LLC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.