NEUSOFT MEDICAL SYSTEMS CO., LTD.
4 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (4)
- Inspections (3)
Devices registered here (1)
- System, X-Ray, Tomography, ComputedProduct code JAK (opens in a new tab) · Class II · Radiology · 21 CFR 892.1750
Recalls & enforcement
- Class IITerminated2018-10-22Software defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal s…
- Class IITerminated2017-05-19Software defect
- Class IITerminated2016-08-31The Firm discovered during contrast agent tracking scan, when the expected concentration of contrast agent is reached, later planned scan may be interrupted. The system must be powered down and restarted to clear the interruption error before scanning can resume.
- Class IITerminated2015-05-02It was found by R&D in April, 2014 that there was a defect in NeuViz 64 system software version 1.0.5+P09. The icons of side decubitus didn't meet the patient position description, when the scanner position was set to "Right" in the system setting - scanner options.
FDA inspections (3)
- NAI2019-01-24No Action Indicated · Compliance: Devices; Postmarket Assurance: Devices; Radiation Control and Health Safety Act
- VAI2016-02-04Voluntary Action Indicated · Compliance: Devices; Radiation Control and Health Safety Act; Postmarket Assurance: Devices · citations posted
- VAI2016-01-29Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices; Radiation Control and Health Safety Act · citations posted
Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “NEUSOFT MEDICAL SYSTEMS CO., LTD. (FEI 3003575350) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3003575350-neusoft-medical-systems-co-ltd/
Alerts for NEUSOFT MEDICAL SYSTEMS CO., LTD.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.