VILLA SISTEMI MEDICALI S.P.A.
2 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (2)
Devices registered here (3)
- Device, Spot-FilmProduct code IXL (opens in a new tab) · Class II · Radiology · 21 CFR 892.1670
- Device, Beam Limiting, X-Ray, DiagnosticProduct code KPW (opens in a new tab) · Class II · Radiology · 21 CFR 892.1610
- Table, RadiologicProduct code KXJ (opens in a new tab) · Class II · Radiology · 21 CFR 892.1980
Recalls & enforcement
- Class IIOngoing2019-11-04An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail bef…
- Class IITerminated2016-12-27Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “VILLA SISTEMI MEDICALI S.P.A. (FEI 3002808461) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3002808461-villa-sistemi-medicali-s-p-a/
Alerts for VILLA SISTEMI MEDICALI S.P.A.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.