RegWardenFDA facility compliance records
Facility

STAAR SURGICAL AG

5 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2013: 1 recalls; 2016: 1 recalls; 2023: 1 recalls; 2026: 2 recalls1’13’14’151’16’17’18’19’20’21’221’23’24’252’26
  • Recalls (5)

Devices registered here (2)

Recalls & enforcement

  • Class IIOngoing2026-04-20
    Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
  • Class IIOngoing2026-01-21
    Toric Implantable Collamer for correction/reduction of myopia in phakic eyes with astigmatism were labeled as 12.1 mm length but are 12.6 mm length lenses, so there is the potential for excessive vault, which may cause: Significant reduction of irido-corneal angles, Significant s…
  • Class IIOngoing2023-09-29
    Their is a potential that intraocular implant devices may not meet specifications.
  • Class IITerminated2016-05-02
    The STAAR AQ Cartridge-FP may malfunction and may crack during loading or delivery of the IOL (Intraocular Lenses).
  • Class IITerminated2013-06-10
    STAAR Surgical Company is recalling a limited number of nanoFLEX CC4204A and Afinity CQ2015A intraocular lenses due to the potential presence of glass particles in the storage vial.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “STAAR SURGICAL AG (FEI 3002808259) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3002808259-staar-surgical-ag/

Not open yet

Alerts for STAAR SURGICAL AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.