RegWardenFDA facility compliance records
Facility

FHC, INC.

3 recall(s) on record — most severe on record: Class I.

Activity by year

FDA actions on record by year2018: 1 recalls; 2019: 1 inspections; 2020: 1 recalls; 2024: 1 recalls; 2025: 1 inspections201712018120191202020212022202312024120252026
  • Recalls (3)
  • Inspections (2)

Devices registered here (1)

  • Stylet, Surgical, General & Plastic Surgery
    Product code GAH (opens in a new tab) · Class I · General, Plastic Surgery · 21 CFR 878.4800

Recalls & enforcement

  • Class IOngoing2024-10-04
    A complaint was received (MDR 3002250546-2024-00003) prompting an investigation. The investigation revealed a schematic for a spacer tube which is too long relative to the size of other components was included in the build order for Lot 244517 of P/N 66-IT-06P. Use of the product…
  • Class IITerminated2020-06-17
    An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.
  • Class IITerminated2018-05-10
    The FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove …

FDA inspections (2)

  • VAI2025-07-10
    Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted
  • VAI2019-07-24
    Voluntary Action Indicated · Compliance: Devices; Postmarket Assurance: Devices · citations posted

Source: U.S. FDA Inspections Dashboard (opens in a new tab) (device inspection classifications, 2016–present). Data as of 2026-10-06. The dashboard profile can take several seconds to load.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “FHC, INC. (FEI 3002250546) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3002250546-fhc-inc/

Not open yet

Alerts for FHC, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned