RegWardenFDA facility compliance records
Facility

FCI OPHTHALMICS, INC.

16 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2012: 1 recalls; 2013: 2 recalls; 2014: 1 recalls; 2015: 2 recalls; 2017: 1 recalls; 2018: 3 recalls; 2020: 2 recalls; 2021: 1 recalls; 2022: 2 recalls; 2026: 1 recalls1’122’131’142’15’161’173’18’192’201’212’22’23’24’251’26
  • Recalls (16)

Devices registered here (1)

Recalls & enforcement

20261 record
20222 records
20211 record
  • Class IIOngoing2021-07-08
    Due to failure to acquire pre-market clearance for its high resolution tomographic/biomicroscope imaging device (Model 6000) prior to commercializing the device.
20202 records
  • Class IITerminated2020-12-07
    When using software 1.90.2.09 or 1.90.8.06 and using modality worklist functionality for patient data transfer, the selection of the patient in the patient list may not match the patient information displayed on the right side of the screen
  • Class IITerminated2020-08-06
    Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potent…
20183 records
  • Class IITerminated2018-12-06
    VisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occu…
  • Class IITerminated2018-10-16
    Under certain fault conditions, the existing design may expose the operator to a hazardous voltage when contacting the metal enclosure (including mounting screws) around the up/down switch.
  • Class IITerminated2018-06-25
    Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
20171 record
20152 records
  • Class IITerminated2015-10-27
    IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
  • Class IITerminated2015-01-22
    Software defect in the FORUM Viewer versions 3.1 and 3.2 which may lead to misinterpretation of the optical coherence tomography (OCT) data.
20141 record
20132 records
  • Class IITerminated2013-10-23
    The user database delivered for download contained an erroneous value. Specifically the "aO" constant of the lens AMO Sensar AR40E is listed as "0.0" in the internal lens database of the IOLMaster. The correct value is -2.420. This could lead to the implantation of lens with a wr…
  • Class IITerminated2013-04-18
    Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.
20121 record
  • Class IITerminated2012-10-25
    A manufacturing defect has been identified which could results in the small tube becoming dislodged between the luer fitting and the check valve.

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)

Cite this record

RegWarden. “FCI OPHTHALMICS, INC. (FEI 3000210020) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/3000210020-fci-ophthalmics-inc/

Not open yet

Alerts for FCI OPHTHALMICS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.