RegWardenFDA facility compliance records
Facility

Agiliti Health - Hays

1 recall(s) on record — most severe on record: Class II.

Activity by year

FDA actions on record by year2024: 1 recalls20172018201920202021202220231202420252026
  • Recalls (1)

Devices registered here (1)

Recalls & enforcement

  • Class IIOngoing2024-01-25
    Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a…

Related

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)