Skytron, LLC
1 recall(s) on record — most severe on record: Class II.
Activity by year
- Recalls (1)
- Warning letters (1)
Devices registered here (1)
- Apparatus, Exhaust, SurgicalProduct code FYD (opens in a new tab) · Class II · General, Plastic Surgery · 21 CFR 878.5070
Recalls & enforcement
- Class IIOngoing2023-08-01When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and t…
FDA warning letters (1)
- Warning Letter2026-05-19Warning Letter (opens in a new tab) · Grand Rapids Michigan
Source: U.S. FDA Compliance Actions Dashboard (opens in a new tab) (device warning letters, 2016–present). Data as of 2026-10-06. Each link opens an FDA search for the letter; FDA can take several weeks to post a letter, so a very recent one may not appear yet.
Related
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this facility record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Skytron, LLC (FEI 1825014) — FDA facility record.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/facility/1825014-skytron-llc/
Alerts for Skytron, LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.