ULRICH MEDICAL USA, INC.
3 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (3)
Registered facilities (1)
- ULRICH MEDICAL USA, INC. FEI 3005532635 · TX
Recalls & enforcement
- Class IITerminated2019-12-31Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
- Class IITerminated2019-03-08Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.
- Class IITerminated2017-03-07Surgical technique was revised to prevent static plate screws blocking mechanisms from disassociating.