PTW Freiburg
4 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (4)
Registered facilities (1)
- PTW FREIBURG FEI 3007360740 · DE
Recalls & enforcement
- Class IIOngoing2026-08-04When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
- Class IIOngoing2024-02-08Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation ca…
- Class IIOngoing2023-06-07Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration…
- Class IITerminated2015-05-27In some cases, especially with high dose-per-pulse-rates typical for particle (specifically proton) therapy, measuring errors can result. Individual measurement intervals can be lost that are not visible to the user when there is a high dose-per-pulse rate resulting in a too-low …
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “PTW Freiburg — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-9010117-ptw-freiburg/
Alerts for PTW Freiburg
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.