THE DAAVLIN DISTRIBUTING CO.
3 recall(s) on record — most severe on record: Class II.
Recalls by year
- Recalls (3)
Registered facilities (1)
- THE DAAVLIN DISTRIBUTING CO. FEI 1526255 · OH
Recalls & enforcement
- Class IIOngoing2026-03-05Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Class IIOngoing2025-12-10Some of the Calibration/Output Certificates had the outputs reversed.
- Class IIOngoing2022-04-26Software issue, resuming an interrupted treatment will result in swap of utilized calibration treatment distance 0" inches to "9. Resulting in over or under dosing.
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. We restate the public record only — measured facts, never risk verdicts.
Cite this record
RegWarden. “THE DAAVLIN DISTRIBUTING CO. — FDA compliance profile.” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/company/oo-1526255-the-daavlin-distributing-co/
Alerts for THE DAAVLIN DISTRIBUTING CO.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.